Victor 14 Process Calibratior
]
type:
Victor 14 Process Calibratior
brand:Victor
model:
Victor 14
supplier:
Xing Chuang Li Gao Instrument & Meter
brief product description:
- Voltage accuracy to ±0.002% reading ± 0.02% range with the max. output current 5mA.
- When 1mA exciting current is presented, the resistance output range up to 400 Ω
- Measurement: Thermocouple, Thermo resistance,
specification
Specification
Output | Range | Output range | Resolution | Accuracy | Explanation |
Voltage | 100mV | -10.000~110.000mV | 0.001mV | ±0.02% reading ± 0.02% range | The max. output current 5mA |
1V | -0.10000~1.10000V | 0.01mV |
Resistance | 400Ω | 0.00~400.00Ω | 0.01Ω | ±0.02% reading ±0.02% range | 1mA exciting current (note1.2) |
Thermocouple | R | -40~1760 ℃ | 1℃ | -40~100℃: 1.5℃ | Adopt the ITS-90 temperature standard (Note 3) |
100~1760℃: 1.2℃ |
S | -20~1760℃ | 1℃ | -20~100℃: ±1.5℃ |
100~1760℃: ±1.2℃ |
B | 400~1800℃ | 1℃ | 400~600℃: ±2.0℃ |
600~800℃: ±1.5℃ |
800~1800℃: ±1.1℃ |
E | -200~1000 ℃ | 0.1℃ | -200~-100℃: ±0.6℃ |
-100~600℃: ±0.5℃ |
600~1000℃: ±0.6℃ |
K | -200~1370 ℃ | 0.1℃ | -200~-100℃: ±0.6℃ |
-100~400℃: ±0.5℃ |
400~1200℃: ±0.7℃ |
1200~1370℃: ±0.9℃ |
J | -200~1200 ℃ | 0.1℃ | -200~-100℃: ±0.6℃ |
-100~800℃: ±0.5℃ |
800~1200℃: ±0.7℃ |
T | -200~400 ℃ | 0.1℃ | -200~400℃: ±0.6℃ |
N | -200~1300 ℃ | 0.1℃ | -200~-100℃: ±1.0℃ |
-100~900℃: ±0.7℃ |
900~1300℃: ±0.8℃ |
Thermo-resistance | Pt100 | -200~850℃ | 0.1℃ | -200~0℃ : ±0.3℃ | Adopt the PT100-385 1mA(Note 1.2) |
0~400℃ : ±0.5℃ |
400~850℃ : ±0.8℃ |
Cu50 | -50~150℃ | 0.1℃ | -50~150℃ : ±0.6℃ |
Leave a message for product